EP 8.0 monograph - impurities test for piracetam using a C18 HPLC column

Instrument Type: HPLC

The Thermo Scientific Dionex UltiMate 3000 LC system is applied for the analysis of piracetam and its impurities. The separation was performed on a Thermo Scientific Syncronis C18 HPLC column using the method described in the USP monograph. The results obtained for resolution and asymmetry exceeded the criteria stated in the USP. Resolution between piracetam and impurity A was 4.8 (USP criteria of not less than 3.0). Asymmetry of piracetam was 1.2 (USP criteria of not more than 2.0).

Related substances analysis of piracetam tablets according to the Chinese Pharmacopoeia using a Thermo Scientific Acclaim 120 C18 HPLC column

Instrument Type: HPLC

In this work the related substances analysis of piracetam tablets is performed according to the Chinese Pharmacopoeia. The separation is performed using a Thermo Scientific Acclaim 120 C18 HPLC column with UV detection at 210 nm.

Piracetam tablets assay according to the Chinese Pharmacopoeia using a Thermo Scientific Acclaim 120 C18 HPLC column

Instrument Type: HPLC

In this work the assay of piracetam tablets according to the Chinese Pharmacopoeia is performed. The separation is performed using a Thermo Scientific Acclaim 120 C18 HPLC column with UV detection at 210 nm.