AU197: Anion Determinations in Municipal Wastewater Samples Using EPA Method 300.1 (A) on an Integrated Ion Chromatography System

Instrument Type: IC

Inorganic anions were determined in three municipal wastewater samples according to U.S. EPA Method 300.1 (A). Nine anions were separated on IonPac AS22 anion-exchange column using manually prepared carbonate-bicarbonate eluent. The anions were detected by suppressed conductivity using chemical regenerant mode with the Thermo Scientific Dionex MMS 300 suppressor and 25 mN sulfuric acid at 7 mL/min.