Find methods for your needs

Refine by Feature

No filters available

Displaying 1-1 of 1 results for Tag: Cefepime dihydrochloride

Items per page:
Sort by

Related substances analysis of cefepime dihydrochloride hydrate using a C18 HPLC column as per JP XVI and KP X monograph method

Instrument Type: HPLC

The Thermo Scientific Dionex UltiMate 3000 LC system is applied for the related substances analysis of cefepime dihydrochloride hydrate. The analysis was performed on a Thermo Scientific Syncronis aQ HPLC column using the method described in the JP XVI and KP X. The results obtained for theoretical plates and %RSD for cefepime met the criteria stated in the JP/KP monograph. The results obtained for theoretical plates of cefepime was 54,132 (JP/KP criteria of not less than 6,000) and %RSD for cefepime was 0.1% (JP/KP criteria of not more than 2).