Find methods for your needs

Dionex IonPac AS20 Column Performance Test Using its QAR Method Be the first to rate this application

Description

Before running any samples, Thermo Scientific recommends that you first confirm the performance of the column by reproducing the lot validation report chromatogram shipped with column. Compare your results with the one reported in the quality assurance report. This record provides an eWorkflow for executing the QAR method on an HPIC system. See column manual for detail.
Market: Environmental; Electronics & Semiconductors; Forensics; Pharma; Chemical
Keywords: Anion, Dionex AERS 500 Suppressor, Ion Chromatography (IC), Water, HPIC, Forensic, environmental, inorganic anions, RFIC, IC, IonPac AS20 column, suppressed conductivity detection, Dionex EGC 500 KOH, QAR, Quality Assurance Report
Matrix: Water
Author: Hua Yang and Jeffrey Rohrer
Affiliation: Thermo Fisher Scientific
Uploaded on 5/9/2018.